  1. ARTICLE 3Interference between dental electricaldevices and pacemakers or defibrillatorsResults from a prospective clinical studyClaude S. Elayi, MD; Stephanie Lusher, RN, MSN,ARNP, BC; Jillian L. Meeks Nyquist, DMD;Yousef Darrat, MD; Gustavo X. Morales, MD;Craig S. Miller, DMD, MSCardiac implantable electronic devices(CIEDs) are indicated to prolong sur-vival in patients with a variety of cardiacdiseases. CIEDs include pacemakers andimplantable cardiac defibrillators (ICDs). Thesedevices are used to treat slow heartbeat (pacing)and to terminate life-threatening arrhythmias bydelivering a large amount of energy (that is,“shock”), respectively. CIED pulse generatorsdetect and interpret internal cardiac signals toprovide effective treatment. Unfortunately, manyexternal electrical devices can mimic or obscureintracardiac signals (that is, electrograms) andpotentially disrupt normal CIED function.1-4Those noncardiac electrical signals, also calledelectromagnetic interference (EMI), may causethe CIED to inhibit pacing, pace irregularly, orpace inappropriately owing to a lack of appro-priate sensing of the heart’s own rhythm. Inter-ference is of concern in thefield of dentistry, inwhich clinicians operate numerous electrical de-vices on a daily basis. Researchers have attemptedto identify dental electrical devices that caninterfere with CIEDs; however, the exact risk forinterference is not well defined in the dentalsetting.5-8The results of several in vitro studies whoseinvestigators used physiological saline baths andbench-top settings suggest possible interferenceThis article has an accompanying online continuing educationactivity available at: http://jada.ada.org/ce/home.Copyrightª2015 American Dental Association. All rightsreserved.ABSTRACTBackground.The authors aimed to determine whether electricaldental devices would interfere with the function of cardiac pace-makers or implantable cardioverter defibrillators (ICDs) inhumans.Methods.The authors exposed asymptomatic nonpacemaker-dependent patients to commonly used electrical dental equipment(for example, battery-operated curing lights, ultrasonic baths, ultra-sonic scalers, electric pulp testers, and electric toothbrushes) in anoutpatient cardiology clinic. The authors operated dental devices atvarious distances and programmed cardiac devices to sense and pace.The authors obtained cardiac tracings using a cardiac programmingunit and a cardiac provider who noted any interference interpretedthe results in real time.Results.The authors enrolled 32 consecutive patients and tested12 pacemakers and 20 ICDs. They did not observe any significantclinical interference in sensing and pacing functions in any patient;however, they noted minor interference without clinical impact inthe telemetry from the cardiac programming unit during use of theultrasonic scaler and bath.Conclusions.Thefindings of this prospective study suggest thatelectrical devices commonly used in dental practices do not inter-fere with the sensing and pacing of contemporary cardiac patients’pacemakers or ICDs. However, they do interfere with the telemetryfrom the cardiac programming unit, without any clinical impact onpatient safety. Thesefindings should help in the development ofclinical guidelines regarding dental management of patients withpacemakers or ICDs.Practical Implications.Electrical dental devices (for example,ultrasonic baths, ultrasonic scalers) induced minor interferencewith programmers that interrogate cardiac devices implanted inpatients; however, overall, dental devices do not appear to interferewith pacemakers’and defibrillators’pacing and sensing function.Key Words.Dental equipment; defibrillators, implantable;electromagneticfields; equipment safety, humans; pacemaker,artificial; telemetry.JADA 2015:146(2):121-128http://dx.doi.org/10.1016/j.adaj.2014.11.016ORIGINAL CONTRIBUTIONSJADA 146(2)http://jada.ada.orgFebruary 2015121
  2. between select dental devices and CIEDs.9-14As a result,statements cautioning dentists to avoid the use of ul-trasonic scalers and other electrical dental devices withpatients who have CIEDs have appeared in the litera-ture.6,15-17However, clinical data are limited. Publishedstudies to date have included a low number of patientsand case reports that have provided conflicting re-sults.18,19In addition, interaction between CIEDs anddental equipment potentially may range from benigninterference (that is, with the CIED’s interrogation unit)to life-threatening inhibition of electrical therapy.Despite major improvements in shielding and EMIrecognition by CIEDs over the last2decades, legitimateconcerns still remain in light of scarce and inconsistentdata.Therefore, we conducted a clinical prospective studyto define whether electrical interference occurs inmodern CIEDs during operation of select dental devicesin close proximity to patients. Furthermore, we aimedto define whether EMI, if any, impeded device function,produced a significant clinical impact (major or minor),or both.METHODSStudy cohort.Between2011and2013, we prospectivelyrecruited all study participants from the implantablecardiac device clinic at our institution. We enrolledconsecutive patients with previously implanted cardiacdevices during their regularly scheduled follow-up visits.Patients aged18to90years of any race or sex who hadpacemakers or ICDs that were functioning appropri-ately—whether they had1,2, or3leads (that is, abiventricular device, allowing cardiac pacing from theright and left ventricles simultaneously)—were eligiblefor enrollment. We excluded patients if they weredependent on a pacemaker (that is, if the patient wassymptomatic without pacemaker function), had anysignificant abnormality in1of the leads that wouldrequire a future intervention, were deemed to be at highrisk for experiencing recurrent ventricular arrhythmia(that is, having had at least2episodes treated by thedevice since the previous interrogation), had a func-tional atrial lead with an ongoing atrial arrhythmiapreventing atrial pacing capabilities (such as atrialfibrillation orflutter), or previously had experienced anytype of EMI. All participants signed informed consentbefore we conducted testing. The University of Ken-tucky Institutional Review Board approved the studyprotocol.Study protocol.Upon enrollment and after receivingan explanation of the study design and signinginformed consent, participants were exposed succes-sively to the following electrical dental devices: Sonicarebattery-operated, electric toothbrush (Philips), Quan-trex Q140H Ultrasonic Cleaning System (L&R Ultra-sonics), Cavitron Select SPS Ultrasonic Scaler(Dentsply), #D-20845Digitest battery-operated electricpulp tester (Parkell), and the Optilux380electric-operated curing light (Kerr). A practicing dentist(J.L.M.N.) individually tested these devices to simulateregular clinical use by turning the devices on and offand operating them at all power levels at varying dis-tances (ranging from0to18inches) from the patient’scardiac device to simulate routine use in the dental of-fice. While dental devices were brought close to thecardiac pulse generator typically located in the leftprepectoral area, we typically scanned the left sideof the thorax by passing the dental device as closeas possible to the cardiac leads. We turned the electronicdental devices to the“on”mode for60seconds, butthen immediately turned them off if the patient expe-rienced symptoms or if the CIED noted significantinterference. We tested each dental device twice,using the following3different pacing and sensingmodes for the CIED: intrinsic patient rhythm, oftensensing the atrium and ventricular signal in ICDs;dual chamber (DDD) sensing and pacing for allleads with a dual inhibited mode at40beats perminute with a very short atrioventricular (AV) delay(80milliseconds) to ensure appropriate atrial sensingand ventricular pacing; and DDD at100bpm withthe longest AV delay based on the device capability(350or400ms)to ensure appropriate atrial pacingand ventricular sensing.A certified electrophysiology nurse practitioner (S.L.)recorded intracardiac electrograms by using the devicecompany’s programmer, which allows interrogation andchanges in the CIED function. The4programmers usedwere Merlin Patient Care System3560(St. Jude Medi-cal), Zoom Latitude Programmer Maintenance, Model3120(Boston Scientific), Renamic4708A (Biotronik),and Medtronic CareLink Programmer, Model2090(Medtronic), respectively. To operate the CIEDs, weconnected the programmers to the CIEDs usingcontinuous inductive telemetry (that is, placing a telem-etry wand on the patient’s clothing over the CIED) orusing radiofrequency (RF) telemetry (that is, a wirelesssystem through an RF antenna). In addition, we recordeda real-time“surface”electrocardiogram (ECG) eitherthrough the respective programmer or, if necessary,through a standard12-lead ECG. We continuouslyABBREVIATION KEY.AP:Atrial pacing.AS:Atrial sensing.AV:Atrioventricular.BiV:Biventricular.BSC:Boston Scien-tific.BTK:Biotronik.CIED:Cardiac implantable electronicdevice.DC:Dual chamber.DDD:Sensing and pacing for allleads with a dual inhibited mode.ECG:Electrocardiogram.EMI:Electromagnetic interference.F:Female.ICD:Implant-able cardiac defibrillator.LVP:Left ventricular pacing.M:Male.MDT:Medtronic.PM:Pacemaker.RF:Radiofrequency.RVP:Right ventricular pacing.SC:Single chamber.SJM:St.Jude Medical.TN:Telemetry noise.VS:Ventricular sensing.ORIGINAL CONTRIBUTIONS122JADA 146(2)http://jada.ada.orgFebruary 2015
  3. monitored and printedintracardiac electrogramsand surface ECGs duringtesting. Then, a specializedcardiologist (electrophysi-ologist) (S.L.) interpretedthefindings, noting anyEMI or abnormalities inpacing. Once the studyprotocol was completed,we thoroughly evaluatedparticipants’health andcardiac devices before dis-missing the patient. Po-tential EMI eventsbetween the CIEDs andthe dental devices wereclassified as major if therewas any effect on theCIED’s function (that is,sensing and pacing func-tion that was not observedduring normal deviceoperation) or a significantclinical impact, and minorif there was any type ofEMI not affecting the de-vice function in combina-tion with no significantclinical impact (thatis, with the CIED pro-grammer). Finally, wenoted artifacts on the reg-ular standard12lead ECG,but did not consider theseto be EMI events becausethey did not relate directlywith CIED function.Data analysis.Werecorded in a MicrosoftExcel spreadsheet andanalyzed the followingdata: age, sex, race, type ofimplanted cardiac device(that is, manufacturer,number and position ofleads, RF capability), timesince implant, type ofEMI, and patient symp-toms, if any. Continuousvariables are reported as amean (standard deviation[SD]) and categoricalvariables as a percentage.TABLECharacteristics of the study population, cardiac devices,and interferences.NO. AGE,YSEX DEVICETYPEDEVICE COMPANY(MODEL)WIRELESSCAPABILITY?TIMEFROMIMPLANT(DAYS)INTERFERENCETYPE*1 2 3 45149 M†PM‡(DC§) SJM¶(Zephyr)No 2,290þ#þ-**- -278 F††PM (DC) MDT‡‡(Adapta)No643þ þ- --355 F PM (DC) BSC§§(H120)No 1,086þ þ- --471 F PM (DC) MDT (ADDRS1)No 1,032þ þ- --557 M PM (DC) BTK¶¶(Cyclos DR)No 1,452þ þ- --656 M ICD##(BiV***)SJM (Unify)Yes11þ þ- --756 M ICD (DC) SJM (Fortify)Yes 1,004þ þ- --863 M ICD (DC) MDT (Virtuoso)Yes 1,907þ þ- --962 M ICD (DC) SJM (Fortify)Yes8þ þ- --1057 F ICD (DC) BSC (T125)No 1,520þ þ- --1167 M ICD (SC†††) SJM (CD1231)Yes666þ þ- --1288 F ICD (DC) SJM (CD2231)No726þ þ- --1353 F ICD (SC) BSC (Teligen)Yes7þ þ- --1462 M ICD (BiV) BSC (H120)Yes 2,315þ þ- --1532 F PM (DC) SJM (Accent)Yes793þ þ- --1654 M ICD (DC) SJM (CD2231)Yes10þ þ- --1740 M PM (DC) SJM (Victory XL)No 2,812þ þ-þ-1862 F PM (DC) MDT (Adapta)No437þ þ- --1938 M ICD (DC) SJM (Fortify)Yes509þ þ- --2056 M PM (DC) MDT (Maximo DR7278)No 2,284þ þ- --2145 M ICD (DC) BSC (Oscar)Yes 1,611þ þ- --2264 M ICD (BiV) MDT (Concerto)Yes 1,977þ þ- --2332 M PM (DC) BTK (Cyclos DR)No14þ þ- --2475 F ICD (BiV) MDT (Concerto)Yes 1,805þ þ- --2569 M ICD (DC) BSC (Teligen)Yes204þ þ- --2644 F ICD (DC) SJM (Ellipse)Yes378þ þ- --2763 M ICD (BiV) BSC (Energen)Yes8þ þ- --2860 M ICD (BiV) BTK (383547)No 2,323þ þ- --2965 F PM (DC) BSC (Altura)No321þ þ- --3069 M PM (DC) SJM (Accent)Yes961þ þ- --3140 M ICD (SC) SJM (Fortify)Yes15þ þ- --3270 M ICD (BiV) BSC (Cognis)Yes10þ þ- --* Interference types include the following: 1¼Cavitron Select SPS Ultrasonic Scaler (Dentsply); 2¼QuantrexQ140H Ultrasonic Cleaning System (L&R Ultrasonics); 3¼Optilux 380 electric-operated curing light (Kerr);4¼Sonicare battery-operated, electric toothbrush (Philips); 5¼#D-20845Digitest battery-operated electricpulp tester (Parkell).†M: Male.‡PM: Pacemaker.§ DC: Dual chamber.¶ SJM: St. Jude Medical.#þ: Interference detected.** -: Interference not detected.††F: Female.‡‡MDT: Medtronic.§§ BSC: Boston Scientific.¶¶ BTK: Biotronik.## ICD: Implantable cardiac defibrillator.*** BiV: Biventricular (with 3 leads).†††SC: Single chamber.ORIGINAL CONTRIBUTIONSJADA 146(2)http://jada.ada.orgFebruary 2015123
  4. RESULTSPatient and device characteristics.Thirty-two consec-utive patients were eligible and enrolled in the study.Thetablelists the characteristics of the study popu-lation, cardiac devices tested, and EMI with thedifferent dental devices tested. All participants wereable to complete the testing protocol withoutcomplication. Twenty-one (65.6%) participants weremale,28(87.5%) were white, and4(12.5%) were Af-rican American. Ages ranged from32to88years(mean57.9[SD¼13.1] years). Twenty individuals(62.5%) had ICDs, and12(37.5%) had pacemakers,with a total of7(25%) having biventricular pacing.Thirteen patients (40.6%) had St. Jude Medical car-diac devices,9(28.1%) had Boston Scientific devices,7(21.9%) had Medtronic devices, and3(9.4%) hadBiotronik pacemakers. All cardiac devices wereimplanted in the left side of the chest with a meantime of973.1[SD¼871.9] days since deviceimplantation.Electromagnetic interference.We did not observemajor EMI in any of the participants; however, wenoted minor EMI events. Those events occurred typi-cally at a distance of less than18inches, with the use ofthe ultrasonic scaler and the ultrasonic cleaning systemin all cases. Examples of minor EMI events duringdental device operation included noise on the cardiacelectrogram channels (Figure1) and complete loss ofcardiac telemetry (Figure2). Finally, we noted somenoise limited to the surface ECG (Figure3). In1participant (patient17), we saw an unexpected ven-tricular pacing activity when the electrophysiologistbrought the sonic toothbrush within12inches of thestudy participant (Figure4). We observed a changefrom atrial and ventricular sense behavior of thepacemaker (Victory XL DR Model5816, St. JudeMedical) to atrial sense and ventricular pace with ashort AV delay that returned to normal after a fewbeats. We immediately removed the dental device, andthe patient remained asymptomatic. We furtherexplored this particular interaction with differentpacing modes and discussed this interaction with thecardiac device company’s technical services represen-tative. We concluded that this particular interactionoccurred because the pacemaker was programmed in atracking mode (DDD) with the autocapture featureturned on. For this device, the complete loss of cardiactelemetry triggered the autocapture threshold search tooccur. This was a minor interaction and illustratednormal device behavior. Furthermore, it could onlyoccur while the device was being interrogated with theprogrammer, which would not occur in a typical dentalsetting.Finally, it is noteworthy that EMI occurred whetherthe programmer was connected to the CIED via theconvention wand placed over the CIED or through awireless remote RF telemetry.DISCUSSIONFor the last several decades, dentists have questionedthe safety of providing dental care to patients whohave CIEDs. Their concern is legitimate as data arescarce and include contradictory results as to whetherelectrical dental equipment interferes with CIEDfunction. In1998, Miller and colleagues9observedthat an electrosurgical unit, ultrasonic bath cleaner,and magnetorestrictive ultrasonic scaler could inhibitin vitro atrial and ventricular pacing within a cardiacpacemaker. In contrast, Garofalo and colleagues10reported in2002that4of5electric apex locatorsdid not interfere with pacemaker function whentested in vitro and concluded that these instrumentscould be used safely on patients. In2007, Brandand colleagues11tested10various electrical dentalFigure 1.Rhythm strip showing intracardiac channel noise during the ultrasonic cleaning system operation. Tracings from top to bottom: Surfaceelectrocardiogram (ECG), atrial channel, ventricular channel, and device markers. Note the noise on the atrial and ventricular channel recognizedon the marker channel as telemetry noise (one of the biventricular implantable cardiac defibrillators [ICDs] from Boston Scientific). The surface ECGon thefirst channel verifies that the ICD works appropriately despite the noise. AP: Atrial pacing. LVP: Left ventricular pacing. RVP: Right ventricularpacing. TN: Telemetry noise.ORIGINAL CONTRIBUTIONS124JADA 146(2)http://jada.ada.orgFebruary 2015
  5. devices on3different types of ICDs in vitro andfound that1ultrasonic bath interfered with2ICDs.In a single patient with a cardiac-implanted pace-maker, study investigators found that an ultrasonicscaler did not alter heart rhythm as monitored by apacemaker technician.19Consistent with this result,in2006Wilson and colleagues18found that electronicapex locators and electric pulp testers did notinterfere with the pacing of cardiac devices whentested in vivo.Figure 2.Rhythm strip showing intermittent complete loss of intracardiac telemetry during the ultrasonic scaler operation. Tracings from top tobottom: Surface electrocardiogram (ECG), device markers, atrial channel, ventricular channel, and farfield. Note the intermittent complete loss ofintracardiac telemetry for all the channels except the surface ECG that verifies proper atrial pacing and ventricular sensing despite the absence oftelemetry (noted in one of the dual-chamber implantable cardiac defibrillators from St. Jude Medical). AP: Atrial pacing. VS: Ventricular sensing.ORIGINAL CONTRIBUTIONSJADA 146(2)http://jada.ada.orgFebruary 2015125
  6. In2010, Roedig and colleagues12studied thein vitro effects of electrical dental equipment on newermodels of pacemakers and ICDs. They found that theultrasonic scaler, ultrasonic bath, and battery-operatedcomposite curing light could interfere with CIED.However, those EMI events may not be clinically sig-nificant as they did not seem to disrupt CIED function,but instead only affected the CIED interrogation unitfunction (also called a programmer). In cardiology andin this study, the programmer is used to access theFigure 3.Rhythm strip showing noise limited to the surface electrocardiogram (ECG) during the ultrasonic scaler operation. Tracings from top tobottom: Surface ECG, device markers, and atrial channel. Note the noise on the surface ECG but not on the intracardiac channel electrograms. AS:Atrial sensing. VS: Ventricular sensing.Figure 4.Example of a complete loss of cardiac telemetry triggering an autocapture threshold search during the ultrasonic scaler operation. Tracingsfrom top to bottom: Surface electrocardiogram (ECG), device markers, atrial channel, and ventricular channel. Note the autocapture threshold searchinitiated by loss of telemetry. Thefirst 2 beats starting at the arrow represent ventricular capture with a short atrioventricular delay (at the pro-grammed output); the next 2 beats are loss of ventricular capture (at an output 0.25 volts less than the initial programmed output) with fusion of thebackup pulse; the next 2 beats regained capture with an increase of the output (by 0.125 volts), which completed the autocapture test. The cardiacdevice then resumed normal function because the dental device was removed. AS: Atrial sensing; VS: Ventricular sensing.ORIGINAL CONTRIBUTIONS126JADA 146(2)http://jada.ada.orgFebruary 2015
  7. CIED’s hard drive via a wand placed on the skin incontact with the CIED or through a wireless antenna.When connected to the CIED, the programmer allowsretrieval of data, change in programming, real-timerecording, and telemetry monitoring. When dentalequipment such as the Cavitron Select SPS UltrasonicScaler is turned on and is in close proximity to theCIED, the programmer loses its telemetry connectionwith the CIED (seeFigure2). Thus, without a simul-taneous real-time recording through the programmerand ECG, it is difficult to be certain whether the CIEDcontinues to pace and sense appropriately.Therefore, we conducted a prospective clinicalstudy to clarify some of the questions raised previ-ously. To our knowledge, this is the largest in vivostudy to investigate EMI of CIED activity during theuse of several commonly used dental devices. Wetested common modern CIEDs produced by4majorcompanies. Our study results show that there wasminor EMI between some dental devices (ultrasonicscalers and ultrasonic cleaning systems) and thedevice programmer. Therefore, noted EMI in previousin vitro studies could be attributed to similar in-teractions. However, there was not a single majorinteraction among the32patients enrolled. CIEDs’sensing and pacing functions were never disruptedwhile dental equipment operated, providing somereassurance for the use of dental equipment withmodern CIEDs.In contrast to the results of previous in vitro studieswhose authors reported EMI, the lack of EMI eventsnoted in humans during this study also could beattributed to the shielding properties of newer models ofCIED. Compared withfirst-generation CIEDs, modernCIED circuitry is sealed hermetically to protect patientsfrom electromagnetic waves to which they are exposedin everyday life. In this study, we also verified propersensing and pacing of CIEDs during the loss of thetelemetry signal on the programmer, because this issuehas been raised previously.20By temporarily“forcing”the CIED to sense and pace, we confirmed the appro-priate CIED pacing and sensing function by having asimultaneous surface ECG recording while the dentaldevice disabled the link between the CIED and theprogrammer unit.Typical of clinical studies, there are limitations toconsider. We did not test every type of CIED thatwas available in clinical practice. However, we tested23different models of modern CIEDs. Furthermore, theCIEDs that we did not test have shielding technologythat is similar to those devices that we did test. There-fore, it is reasonable to extrapolate our data to modernavailable devices. Also, we did not test patients whowere dependent on CIEDs in this study due to safetyconcerns. However, there is no mechanistic reason asto why those results should differ among patients whoare dependent on pacemakers. Finally, the number ofpatients in this study was relatively small.CONCLUSIONSCommonly used electrical dental devices (for exam-ple, ultrasonic cleaning systems, ultrasonic scalers)presented minor EMI with programmers that interro-gate cardiac devices implanted in patients. Those in-teractions are not typically observed in a dental settingbecause cardiac devices would never be interrogatedthere, and the observed interactions were not associatedwith significant clinical implications. Thus, the dentaldevices tested did not appear to interfere with pace-makers’and defibrillators’pacing and sensing functionof newer models of pacemakers and ICDs. Largerstudies are needed to confirm ourfindings.nDr. Elayi is an associate professor, Department of Internal Medicine,Division of Cardiovascular Medicine, Clinical Electrophysiology and Pacing,Gill Heart Institute, University of Kentucky, Lexington, KY, and LexingtonVeterans Administration Medical Center, Lexington, KY.Ms. Lusher is a nurse practitioner, Department of Internal Medicine,Division of Cardiovascular Medicine, Clinical Electrophysiology and Pacing,Gill Heart Institute, University of Kentucky, Lexington, KY, and LexingtonVeterans Administration Medical Center, Lexington, KY.Dr. Meeks Nyquist is a research fellow, Division of Oral Diagnosis, OralMedicine, and Oral Radiology, Department of Oral Health Practice, Collegeof Dentistry, University of Kentucky, Lexington, KY.Dr. Darrat is an assistant professor, Department of Internal Medicine,Division of Cardiovascular Medicine, Clinical Electrophysiology and Pacing,Gill Heart Institute, University of Kentucky, Lexington, KY, and LexingtonVeterans Administration Medical Center, Lexington, KY.Dr. Morales is an assistant professor, Department of Internal Medicine,Division of Cardiovascular Medicine, Clinical Electrophysiology and Pacing,Gill Heart Institute, University of Kentucky, Lexington, KY, and LexingtonVeterans Administration Medical Center, Lexington, KY.Dr. Miller is a professor, Oral Medicine Section, Division of Oral Diag-nosis, Oral Medicine, and Oral Radiology, Department of Oral HealthPractice, MN324, College of Dentistry, University of Kentucky,800Rose St.,Lexington, KY40536-0297, e-mailcmiller@uky.edu. Address correspon-dence to Dr. Miller.Disclosure.None of the authors reported any disclosures.The authors thank Bill Everly of St. Jude Medical Co., Deborah Williamsof Medtronic, and Jill Elmore of Boston Scientific for their technical supportduring this study.1.Niehaus M, Tebbenjohanns J. Electromagnetic interference in patientswith implanted pacemakers or cardioverter-defibrillators.Heart.2001;86(3):246-248.2.Pinski SL, Trohman RG. Interference in implanted cardiac devices,part II.Pacing Clin Electrophysiol.2002;25(10):1496-1509.3.Dyrda K, Khairy P. Implantable rhythm devices and electromagneticinterference: myth or reality?Expert Rev Cardiovasc Ther.2008;6(6):823-832.4.Napp A, Joosten S, Stunder D, et al. Electromagnetic interference withimplantable cardioverter-defibrillators at power frequency: an in vivostudy.Circulation.2014;129(4):441-450.5.Drisko CL, Cochran DL, Blieden T, et al; Research, Science andTherapy Committee of the American Academy of Periodontology. Positionpaper: sonic and ultrasonic scalers in periodontics. Research, Science andTherapy Committee of the American Academy of Periodontology.J Periodontol.2000;71(11):1792-1801.ORIGINAL CONTRIBUTIONSJADA 146(2)http://jada.ada.orgFebruary 2015127
  8. 6.Adams D, Fulford N, Beechy J, MacCarthy J, Stephens M. The cardiacpacemaker and ultrasonic scalers.Br Dent J.1982;152(5):171-173.7.Possible electromagnetic interference with cardiac pacemakers fromdental induction casting machines and electrosurgical devices. Council onDental Materials and Devices.JADA.1973;86(2):426.8.Dawes JC, Mahabir RC, Hillier K, et al. Electrosurgery in patients withpacemakers/implanted cardioverter defibrillators.Ann Plast Surg.2006;57(1):33-36.9.Miller CS, Leonelli FM, Latham E. Selective interference with pace-maker activity by electrical dental devices.Oral Surg Oral Med Oral PatholOral Radiol Endod.1998;85(1):33-36.10.Garofalo RR, Ede EN, Dorn SO, Kuttler S. Effect of electronicapex locators on cardiac pacemaker function.J Endod.2002;28(12):831-833.11.Brand HS, Entjes ML, Nieuw Amerongen AV, van der Hoeff EV,Schrama TA. Interference of electrical dental equipment with implantablecardioverter-defibrillators.Br Dent J.2007;203(10):577-579.12.RoedigJJ, Shah J, Elayi CS, Miller CS. Interference of cardiac pace-maker and implantable cardioverter-defibrillator activity during electronicdental device use.JADA.2010;141(5):521-526.13.Gomez G, Duran-Sindreu F, Jara Clemente F, et al. The effects of sixelectronic apex locators on pacemaker function: an in vitro study.IntEndod J.2013;46(5):399-405.14.Idzahi K, de Cock CC, Shemesh H, Brand HS. Interference of elec-tronic apex locators with implantable cardioverter defibrillators.J Endod.2014;40(2):277-280.15.Trenter SC, Walmsley AD. Ultrasonic dental scaler: associatedhazards.J Clin Periodontol.2003;30(2):95-101.16.Cardiac arrhythmias. In: Little JW, Falace DF, Miller CS, Rhodus NL,eds.Dental Management of the Medically Compromised Patient.8th ed. St.Louis, MO: Elsevier;2013:67-80.17.Stoopler ET, Sia YW, Kuperstein AS. Does ultrasonic dental equip-ment affect cardiovascular implantable electronic devices?J Can DentAssoc.2011;77:b113.18.Wilson BL, Broberg C, Baumgartner JC, Harris C, Kron J. Safety ofelectronic apex locators and pulp testers in patients with implanted cardiacpacemakersor cardioverter/defibrillators.J Endod.2006;32(9):847-852.19.Carlson BK. Pacemakers and dental devices.JADA.2010;141(9):1052-1053.20.Elayi CS, Shah J, Miller CS. Author’s response.JADA.2010;141(9):1053-1054.ORIGINAL CONTRIBUTIONS128JADA 146(2)http://jada.ada.orgFebruary 2015